Multi-Objective Design Optimization of Tibial Inserted Conformity for Improved Wear and Kinematics Performance using Generalized Regression Neural Networks

Chotchuang Phombut1

Supakit Rooppakhun1, Email

Attasit Wiangkham2

Atthaphon Ariyarit1

1School of Mechanical Engineering, Institute of Engineering, Suranaree University of Technology, Thailand 111 University Avenue, Muang, Nakhon Ratchasima, 30000, Thailand
2Department of Industrial Engineering, Faculty of Engineering, Srinakharinwirot University, Ongkharak, Nakhon Nayok, 26120, Thailand

 

Abstract

The design of the tibial insert component significantly influences the performance of knee prostheses, particularly with respect to mechanical wear and kinematics. This study aims to optimize both wear and kinematics performance of tibial insert conformity in total knee arthroplasty (TKA) using Generalized Regression Neural Networks (GRNNs). Finite element analysis (FEA) was conducted based on a central composite design (CCD), with coronal conformity (Cc) and sagittal conformity (Sc) as input variables. GRNN models were developed to predict wear and kinematic responses, achieving high predictive accuracy with R2 and mean absolute percentage error (%MAPE) values of 0.972 and 1.950% for kinematics and 0.995 and 9.220% for wear, respectively. Multi-objective optimization was performed using the non-dominated sorting genetic algorithm II (NSGA-II) combined with GRNNs. The optimal ranges for Cc and Sc that minimized both wear and kinematic deviations were identified. Results revealed that coronal conformity had a greater impact on volumetric wear rate, while sagittal conformity more strongly influenced kinematic behavior. These findings underscore the significance of conformity design in improving the biomechanical performance of total knee arthroplasties (TKA).

Multi-Objective Design Optimization of Tibial Inserted Conformity for Improved Wear and Kinematics Performance using Generalized Regression Neural Networks