Performance Evaluation of Light-Emitting Diode Overhead Phototherapy Devices: Spatial Irradiance Distribution and Effective Irradiated Area at Clinical Distances

Supawadee Tubglam1,4, Email

Phadungsak Rattanadecho2

Sanong Ekgasit3

1Medical Engineering Program, School of Engineering, Thammasat University, Pathum Thani, 12120, Thailand
2Hub of Talents Electromagnetic Energy Utilization in Medical Engineering, Department of Mechanical Engineering, School of Engineering, Thammasat University, Pathum Thani, 12120, Thailand
3Department of Chemistry, Faculty of Science, Chulalongkorn University, Bangkok, 10330, Thailand
4Department of Child and Adolescent Nursing, Faculty of Nursing, Thammasat University, Pathum Thani, 12120, Thailand

 

Abstract

Neonatal hyperbilirubinemia is a common condition that may cause irreversible neurological injury if inadequately treated. Effective phototherapy depends on adequate light delivery, including sufficient spectral irradiance and uniform spatial distribution across the irradiated area. However, data on the irradiance distribution and effective irradiated area of light-emitting diode (LED) overhead phototherapy devices at commonly used clinical distances remain limited. This study evaluated whether three commercial LED overhead phototherapy devices achieved American Academy of Pediatrics (AAP)-referenced irradiance criteria at source-to-surface distances of 20, 25, and 30 cm in low-and high-intensity modes. Spectral irradiance was measured across a 60-point grid using a calibrated radiometer. Irradiance distribution and effective irradiated area were analyzed using three-dimensional surface plots, heat maps, and labeled heat maps. In low-intensity mode, peak irradiance reached 61–62 µW/cm2/nm, but only 30–53.3% of grid locations met the lower bound of the AAP conventional range (≥8 µW/cm2/nm). In high-intensity mode, peak irradiance reached 147.8 µW/cm2/nm, while only 21.7–41.7% met the AAP intensive threshold (≥30 µW/cm2/nm). Irradiance distribution was non-uniform and patchy, with limited, fragmented threshold-level coverage, off-center peaks, and localized hot spots. Systematic spatial irradiance assessment under device-specific operating conditions and manufacturer-guided source-to-surface distances is recommended to support effective and safe phototherapy.

Performance Evaluation of Light-Emitting Diode Overhead Phototherapy Devices: Spatial Irradiance Distribution and Effective Irradiated Area at Clinical Distances