Numerical Analysis of the Effect of Metal Implant Length and 5G Source Position on Specific Absorption Rate and Temperature Distribution in Human Head

Deepshikha Bhargava1, Email

Phadungsak Rattanadecho2

1Department of Mechanical Engineering, Faculty of Engineering, Kasetsart University, Bangkok, 10900, Thailand
2Department of Mechanical Engineering, Faculty of Engineering Thammasat University, Pathum Thani, 12120, Thailand

 

Abstract

The rapid development of 5G networks has raised concerns about how higher-frequency signals interact with the human body, particularly in individuals with metallic implants. In this study, a two-dimensional heterogeneous head model is exposed to a 5G source operating at 3.6 GHz and 4.4 GHz with 1 W radiated power. A titanium pin is inserted in the bone layer of the model to represent a cranial implant. The finite element method (FEM) is used to analyze energy absorption by solving a coupled electromagnetic (EM) and bioheat transfer model. The effects of varying the implant length and the distance between the 5G source and the head on the electric field, Specific Absorption Rate (SAR), and temperature distribution in the head are examined. The results show that certain combinations of source distance and pin length lead to higher absorption due to stronger current coupling. The hotspot location within the head also shifts based on the source distance and implant length. The highest SAR obtained is 21.58 (W/kg), and the maximum temperature rise is 0.53 (°C). While SAR exceeds recommended safety limits, temperature increases remain below harmful thresholds.

Numerical Analysis of the Effect of Metal Implant Length and 5G Source Position on Specific Absorption Rate and Temperature Distribution in Human Head