Optimization and Validation of 4π Ionization Chamber for Precise I-131 Half-Life Measurement and Reduced Uncertainty 

Chailoek Tanghengjaroen1

Dhonluck Manop1

Yongyut Manjit1

Worasit Palkawong Na Ayuthaya2

Tanapol Dachviriyakij3

Narudom Noulkhow4

Santi Raksawong5

Panya Khaenamkaew1, Email

1Department of Basic Science and Physical Education, Faculty of Science at Sriracha, Kasetsart University, Sriracha, Chonburi, 20230, Thailand
2Department of Resources and Environment, Faculty of Science at Sriracha, Kasetsart University, Sriracha, Chonburi, 20230, Thailand
3Ionizing Radiation Metrology Laboratory, Regulatory Technical Support Division, Office of Atoms for Peace, Ministry of Higher Education, Science, Research and Innovation, Bangkok, 10900, Thailand
4Department of Policy and Strategy, National Institute of Metrology (Thailand), Ministry of Higher Education, Science, Research and Innovation, Phatum Thani, 12120, Thailand
5Division of Physics, Faculty of Science and Technology, Muban Chombueng Rajabhat University, Chom Bueng, Ratchaburi, 70150, Thailand

Abstract

Accurate and efficient secondary calibration of radionuclides remains a challenge in radionuclide metrology, with limited guidance on optimizing operational parameters for 4π ionization chambers (4π-ICs). This study addresses this gap by investigating the optimization of key measurement parameters—time and source-to-detector distance—in order to improve the accuracy and efficiency of 4π-ICs used in nuclear medicine and standardization laboratories. The stability of the chamber was assessed using ionization currents from a reference source of holmium-166m (166mHo). Linearity testing was conducted with three iodine-131 (131I) sources at varying activity levels. Measurements were performed over durations ranging from 10 to 80 seconds and distances from 3.00 to 7.00 cm, with standard conditions set at 40 s and 4.00 cm. The measured half-life of 131I was 8.0225 ± 0.0004 days, consistent with published values, and the uncertainty averaged ±0.024% (k = 1). A distance of 3.50 cm produced results similar to 4.00 cm but with reduced uncertainty. The optimal measurement condition was determined to be 35 s at 4.50 cm. This work offers a novel, statistically validated approach for optimizing 4π-IC use, enhancing radiation safety, reducing radioactive material consumption, and improving productivity in radioisotope laboratories.

Optimization and Validation of 4π Ionization Chamber for Precise I-131 Half-Life Measurement and Reduced Uncertainty